HIPP-3

General Data

Pressure < 5×10-10 mbar, Up to 30 mbar, verified for N2
Baking temperature: 120 ºC
Analyzer radius: 200 mm
Mounting flange: NW 200 CF, fixed
Working distance: 0.3 - 9 mm, depending on aperture
Slits: 9
Detector type: MCP/CCD camera
Detector interface: Ø 40 mm MCP
Energy channels: > 1000 simultaneous
Angular channels: > 750 simultaneous
Lens modes: Transmission, Swift Acceleration, Angular, Spatial
Acquisition modes: Swept, fixed
Detector modes: Pulsed, ADC
Intensity deflectors: Yes (x, y)
ISS: Option
Analyser pumping recommendations:
First stage: 2x300 l/s
Second stage: 1x300 l/s
Third stage: 1x300 l/s

Top

High Voltage Electronics Technical Data

Temperature stability: < 2 ppm/ºC (R-version), (typical 0.5 ppm/ºC)
Noise (AV at analyzer): < 1 ppm + < 500 μV, (typical 0.5 ppm + < 200 μV)
Drift: < 20 ppm/year, (typical 10 ppm/year)
Electric isolation: 6 kV
Min. step size: HV100 1.6 mV
Min. step size: DAC 200 μV
DAC: Bits 16
Modular: Yes
Communication: USB

Top

HIPP-3 XPS

Energy resolution: 15 meV at 500 eV
Spatial resolution:
10 μm meV up to 1200 eV
20 μm meV up to 1500 eV


Angular resolved range: ± 9º
Kinetic energy range
Transmission mode 5 - 1 500 eV
Angular mode 5 - 1 500 eV
Spatial mode 5 - 1 500 eV
Pass Energy 5 - 200 eV

Top

HIPP-3 UPS upgrade*

Energy resolution: 1.8 meV at 10 eV
Spatial resolution: 10 μm meV
Angular resolved range: ± 9º

Kinetic energy range
Transmission mode: 2 - 100 eV
Angular mode: 2 - 100 eV
Spatial mode: 2 - 100 eV
Pass Energy: 2 - 10 eV

*Specifications in addition to HiPP-3 XPS

Top

HiPP-3 6 keV upgrade*

Energy resolution: 40 meV at 6 keV
Angular resolved range: ± 9º


Kinetic energy range
Transmission mode: 1 500 - 6 keV
Angular mode: 1 500 - 1950 eV

*Specifications in addition to HiPP-3 XPS

Top
Feedback